Premier series monitor, Quick start – Eagle Microsystems GD-1000 User Manual

Page 2

Advertising
background image

EAGLE MICROSYSTEMS. INC.
Instrumentation for Measurement and Control

2

copyright 2/02

Premier Series Monitor

TABLE OF CONTENTS:

SPECIFICATIONS:

Page 2

QUICK START:

Installing, wiring and display operation.

Page 2

OPERATION:

Main run screen, password entry screen and monitor setup screen.

Page 3

ALARM:

Critical and danger screens.

Page 3/4

CALIBRATION:

Analog input (CH1), Analog output (IOUT).

Page 4/5

BATTERY BACKUP:

Battery status

Page 5

WIRING:

Power, alarms, analog input and outputs

Page 6

SPECIFICATION:

Power Supply

115/230 VAC Switch selectable.

Power Consumption

12W

Display

LCD, backlit, 16 characters by 2 lines, 5mm character height,

5x8 character matrix.

Pushbuttons

Four, tactile dome.

Relays

Three, Mechanical (AC or DC)

Buzzer.

103 DB

Rating 1.2A,

125VAC.

Optional: 5A @ 250VAC.

Analog Input

two, 4 to 20mA.

Analog Output

Isolated 4 to 20mA, 0-500ohm load, active.

Digital Output

20mA Serial (remote computer)

QUICK START

INSTALLATION:

1).

Mount the Premier Series Monitor at eye level. Mount the sensor 12” off the
floor. The Premier Series Monitor is provided with 12’ of cable for the sensor.

2).

Attach the cable from the Premier Series Monitor to the sensor. The
red LED will go on when powered up.

3).

The Premier Series Monitor requires 117VAC (+/-10%) power at 48-62Hz
and 1/4 amp.

After proper installation the monitor is ready to operate

.

DISPLAY:

The top line of the LCD display is formatted to show a selected value along with
its engineering units. The second line normally provides function labels for
the four pushbuttons located below the function. The pushbuttons do not have
any labels.

2

Line1: Value &
units

1

3

4

IHPN 630600

SENSOR

Advertising