Premier series monitor, Calibration – Eagle Microsystems GD-1000 User Manual

Page 4

Advertising
background image

EAGLE MICROSYSTEMS. INC.
Instrumentation for Measurement and Control

4

copyright 2/02

ALARM SCREEN:

Press the pushbutton under CRT or DGR, to configure. Press ESC to return.

1

CRT

Critical alarm screen to set value for relay contact.

2

DGR

Danger alarm screen to set value for relay contact.

3
4

ESC

Go back to the monitor setup screen.

CRITICAL ALARM SETUP SCREEN:

From the critical alarm setup screen you can view the current value. To
change, press the pushbutton under UP or DWN to change the value. Press
the pushbutton under SET to enter the new value and return back to alarm
setup screen.

1

UP

To increase the value.

2

DWN

To decrease the value.

3
4

SET

Go back to the alarm setup screen.

DANGER ALARM SETUP SCREEN:

From the danger alarm setup screen you can view the current value.

To

change, press the pushbutton under UP or DWN to change the value. Press
the pushbutton under SET to enter the new value and return back to alarm
setup screen

1

UP

To increase the value.

2

DWN

To decrease the value.

3
4

SET

Go back to the alarm setup screen.

Premier Series Monitor

CALIBRATION:

SELECT CALIBRATION

From the MONITOR SETUP SCREEN, press the pushbutton under CAL.

1

CAL

Enter calibration screen..

2

ALM

Enter alarm select screen.

3
4

RUN

Go back to the main run screen

CALIBRATION SCREEN:

Press

the pushbutton under CH1 or IOUT

.

Press ESC to return.

1

CH1

Press to calibrate 4-20mA analog input..

2
3

IOUT

Press to calibrate 4-20mA analog output.

4

ESC

Will ask to set new password yes /no

SEL

SO2

CH1 ALM

CRT

DGR

ESC

CRITICAL

10.0

UP

DWN

SET

DANGER

2.0

UP

DWN

SET

MONITOR SETUP

CAL

ALM

RUN

CALIBRATION

CH1

IOUT

ESC