Flintec KA-1 User Manual
Page 3

Junction Box Type KA – Technical Manual, Rev. 1.00 August 2007
Page 3 of 4
I
NTRODUCTION AND TECHNICAL
D
ATA
The Aluminium junction box is designed for the connection of 1 load cells.
Type
No. of load cells
Housing size
Input
Output
KA-1
1
64 x 98 x 34 mm
1x M16
1x M16
The junction box type KA-1 can be connected to the instrumentation with a shielded 6-wire signal cable.
Housing material:
Aluminium
Protection class:
IP66
Cable connection:
By screw terminals
Corner correction:
Not required
M
ECHANICAL
I
NSTALLATION
Look for a mounting location which is more or less dry and protected from environmental stress.
E
LECTRICAL
C
ONNECTIONS
Figure 1: Dimensions in [mm]
L
OAD
C
ELL
C
ABLE
C
ONNECTION
First the cable gland must be loosened. Then you have to feed the load cell cable through the cable gland
unless the shrink tube is fully disappeared in the box.
Afterwards the wires can get connected as follows (see also fig.2)
Cable coulour
Description
Terminal no., example
yellow
= Cable shield
First ground screw
(if applicable, blue)*
= Sense +
1*
green
= Excitation + (Input +)
2*
white
= Signal + (Output +)
3
red
= Signal – (Output –)
4
black
= Excitation – (Input –)
5**
(if applicable, brown)*
= Sense –
6**
* at load cells with 4-wire technology terminal no.1 and no.2 must get connected
** at load cells with 4-wire technology terminal no.5 and no.6 must get connected