Cancelling the timer, 15 setting a high temperature warning alarm – Grant Instruments LTC2 User Manual

Page 17

Advertising
background image

LTC2

30933 V2

Operating Manual

Page 16

www.grantinstruments.com

Note: the LTC2 will continue to control at the set temperature after the timer
reaches zero – the heater will not switch off.

5.2.14 Cancelling the timer

The countdown timer can be easily cancelled.

1. Press the F button twice – the display will show “Cloc”.
2. Press the S button to select.
3. Use the main dial to select “Off” and press the S button to select. The display will

revert to the bath temperature and the timer will be cancelled.

5.2.15 Setting a high temperature warning alarm

The high temperature alarm sets a warning buzzer to sound if the bath temperature
exceeds a defined level.

1. Press the F button three times – the display will show ”Alar”.
2. Press the S button to select.
3. Use the main dial to select “On” and press the S button to select. The display flashes

the high temperature alarm value. Rotate the dial to set the high temperature alarm
required.

4. Press the S button to save and return to liquid temperature.

If the high temperature alarm value is exceeded then LTC2 will sound the buzzer
intermittently and show “-Al-” on the display. The alarm can be acknowledged by
pressing either the F or S buttons. However, the alarm is persistent and will trigger again
if the temperature still exceeds the high temperature value. The alarm can be cancelled
by following the steps above but using the dial to select “Off” in step 3 and pressing the S
button to confirm.

Note: the high temperature alarm will not switch off the heater.

Advertising