0 technical specifications, Operating conditions, Electrical details – Grant Instruments LTC2 User Manual

Page 18: Performance

Advertising
background image

30933 V2

LTC2

Page 17

Operating Manual

www.grantinstruments.com

6.0 Technical specifications


6.1

Operating conditions

Ambient temperature range

5 to 40°C

Altitude above sea level

Up to 2,000m (6,500ft)

Operating environment

Indoor use only

Maximum relative humidity

80% RH up to 31°C decreasing to 50% RH at 40°C


6.2

Electrical details


Mains supply:

220-240V @ 50Hz or 110-120V @ 60Hz

Pollution degree:

2

Installation category:

II


Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.

6.3

Performance

Specification

LTC2

LTC2L

Typical cooling power at an
ambient of 20°C

@ 20°C

250W

@ 0°C

140W

@ -10°C

100W

@ -20°C

35W

Stability (DIN 12876)

±0.1°C †

Uniformity (DIN 12876)

±0.1°C

Settable temperature range

-20°C to 120°C

Working temperature range

-20°C to 100°C

Refrigerant R134a

140gm

Tank capacity

5L

Top opening

110/145mm

Max pump head pressure

210mBar

Max pump flow rate

16L/min

Safety 100°C limit

Cut-out

Safety H

2

O freezing protection

5°C thermostat & switch

Max current consumption (both units)

8.0A

15.5A*

Heater power

1.3kW

1.4kW

*

Requires two separate mains cords

† Temperature stability may be affected (±0.3C) in the presence of strong RF fields (10V/m) at
380-400MHz. This level of interference is very unlikely and is only likely to be encountered in
electrically noisy industrial locations as defined in EN61326)