30933 V2

LTC2

Page 23

Operating Manual

www.grantinstruments.com

Display shows
“Er L”

Incorrect calibration
value

The low temperature calibration point is less
than 40°C below the high calibration point –
choose a lower temperature (see section
5.2.10).

Display shows
“Er d”

Incorrect calibration
value

The calibration value is greater than 10°C
from the set point – check thermometer
reading and set point (see section 5.2.10).

Display shows
“Open”

Faulty temperature
probe

Have a competent person check the probe for
an open circuit fault or contact Grant.

Display shows
“Shrt”

Faulty temperature
probe

Have a competent person check the probe for
a short circuit fault or contact Grant.

Unit showing
erratic
temperatures

Calibration values not
set correctly

Restore the factory calibration settings (see
section 5.2.11) then re-calibrate if required
(see section 5.2.10).

Display shows
“Hot”

Heating water at or
near to boiling without
lid

Heating very large
volumes of liquid with
large thermal losses

Add a lid to reduce thermal losses and leave
unit running with the “Hot” warning present. If
the alarm has not cleared within 1 hour
contact Grant.

Add measures to reduce thermal losses and
leave unit running with the “Hot” warning
present. If the alarm has not cleared within 1
hour contact Grant.

Display shows
“Fuse”

Faulty fuse(s)

Have a competent person check the internal
fuses or contact Grant.

Display shows
“Crct”

PCB fault

Have a competent person check the product
or contact Grant.

Display shows
“Self”

Safety relay fault

Have a competent person check the product
or contact Grant.

Stirrer motor not
rotating

Stirring propeller or
pump impeller is
obstructed

Faulty motor

Clear obstruction.



Have a competent person check the motor or
contact Grant.

Display shows
“Cold”

Accidentally entered
manufacturer reset
mode

Warning can be cleared by pressing S button,
but note that all settings (including any factory
calibration) will be reset to the default values.
Contact Grant Instruments service
department for assistance if required.


For any other errors or service requests, please contact Grant Instruments service
department.