General information – Grant Instruments PCV-3000 User Manual

Page 6

Advertising
background image

2. General Information

Multi-spin PCV-3000 is the next step of development of sms technology: it is a fully
automated device utilising the sms algorithm with good repeatability. Productivity of sms
algorithm using Multi-Spin is 10 times faster (up to 30 min per 1000 samples) than that for
centrifuge-vortex Combi-Spin.

Combi-Spin (the previous name Micro-Spin) was invented and published for the first time by
Biol. Dr. V.Bankovskis in 1989 and after the successful approbation and finishing patented in
1994 (V.K.Bankovskis et al., Riga, Latvia, Pat. No. P94-74).

Spin-mix-spin technology automates the repeated steps of spinning down of the sample
followed by vortex mixing. This cycle of repeats significantly reduces the likelihood of
contamination and therefore errors in PCR analysis as well as the overall time taken for
sample preparation.

PCR reaction preparation times have traditionally been lengthy, the Multi-Spin reduces this
time. Trying to achieve the same sms algorithm manually is around ten times slower (up to
30 min for 100 samples vs 30 min for 1000 samples on the Multi-Spin) and the reproducibility
of results may vary due to slight differences in the vortex and centrifugation stages, a normal
consequence of manual operations.

Page 6

Multi-spin PCV-3000

Operating Instructions

Table 1. Comparison of PCV-4200, PCV-3000 and PCV-6000

Specifications

Combispin

PCV-2400

Multi-Spin

PCV-3000

Multi-Spin

PCV -6000

Speed range max.

2800 rpm

3500 rpm

6000 rpm

RCF max.

700 x g

800 x g

2350 x g

Number of tubes
vortexing

1 individually

12 simultaneously

Time for completing “Spin-

Mix

-Spin” cycle:

for 2 microtubes

60 sec

25 sec

15 sec

for 12 microtubes

5 – 6 min

1 min 30 sec

1 min

for 100 microtubes

60 min

15 min

10 min

Version 2.02 - August 2012

Advertising