6.0 Calibration (GR150 & GP200 only)

6.1.

Performing a single point calibration


A single point calibration can be performed using Labwise. This feature is only available with
the GR150 and GP200. A temperature offset can be applied such that the temperature
displayed by the Optima can be aligned with that taken by an independent temperature
probe.

1. Select Calibration from the main control window menu

The following dialog box will appear.


2. Adjust the ºC Offset such that the Corrected Temp ºC matches the independent

temperature probe reading.
The Current Temp is the temperature measured by the Optima. The Corrected Temp
will be the indicated temperature if the offset is applied.


3. Click the Accept button to apply the offset.