2.1 TRANSDUCER

.

IMPORTANT: All transducers have a directional arrow on the tag

and/or etched into a metal part. Before installing a sensor, please note
proper flow direction. This is critical to sensor operation.

.

IMPORTANT: If you have more than one Rheotherm unit, make sure

the complete serial number of the transducer matches the complete serial
number of the electronics. The transducer and electronics are a matched
set. Components with different serial numbers should not be
interchanged. The transducers have no user serviceable parts, so do not
try to disassemble, as permanent damage may result.

The transducer style supplied with your meter is listed in the model code number in SECTION 6.
Proper installation of the sensor is necessary for achieving accuracy and repeatability. Installation
suggestions for each type of standard transducer are given here. For custom transducer installations,
refer to CUSTOM INFORMATION — SECTION 6.

Be sure wetted surfaces are clean before installing. If cleaning is needed, use non-residue solvent and
wipe dry. If the sensor has a connector box, keep moisture out. Make sure the lid is tightly sealed and,
if supplied, the gasket is in place. Seal conduit lines at the connector box if conduit lines can become
wet.

1.

TU (nonintrusive) — TU

1

'

16

and TU

c transducers, unless they have optional ¼" O.D. ends,

particularly require special care in handling and installing to avoid damage to sensor tube
stubs.

.

CAUTION: TU transducers are made with thin-walled tubing — use

care when installing.

All TU transducers (other than

1

'

16

and

c inch) should have straight line input and output

sections, typically 20 pipe diameters on the inlet and 6 to 10 diameters on the outlet. If
installed vertically, the flow should be going up through the sensor. Connection in the line
is via compression fittings, hose with clamp, threaded fittings or flanges, whichever is
appropriate. Care must be taken not to transmit a twisting force through the transducer's
midsection. The TU transducer, whether flanged or not, must not be used to pull other piping
together or to make up angular mismatch of fittings. The transducer junction box (if supplied)
should never be rotated for any reason.

TU

1

'

16

transducers may be sleeved with a

c" tube for added support. Connection should

always be made to the

1

'

16

" tube, as there is no assured seal between the

1

'

16

" tube and the

sleeve.

Some TU transducers have an integrally mounted cable; do not pull on this cable, or attempt
to remove the fitting where the cable enters the sensor shell.