KERN & SOHN MLS 50-3C User Manual
Page 25
MLS_C-BA-e-1330
25
Ensure that there are no objects on the weighing
plate. Press the
PRINT key.
In the display there appears the value of the
adjustment weight.
Place the required adjustment weight in the centre of
the sample dish, then press the
PRINT key. If the
adjustment was successful, this will be followed by
automatic reset to the menu.
Take away adjustment weight.
The
ESC key may be used to cancel the adjustment.
Errors during adjustment or the use of an incorrect
adjusting weight will result in an error message.
Repeat adjustment.
Return to weighing mode:
Press the
ESC-key repeatedly until the query
”
SAVE?“ appears.
Confirm query by pressing the
PRINT-key or reject it
by pressing the
ESC-key.
Setup
!
Saveing...
Setup
29.01.08
We recommend:
Carry out all adjustments as close as possible to the maximum load of the balance.
Info about test weights can be found on the Internet at: