KERN & SOHN MLS 50-3C User Manual
Page 31
Advertising

MLS_C-BA-e-1330
31
Temperature calibration of third point will take 8 min
after which you will hear an acoustic signal
Enter the temperatur value as indicated on the
external thermometer with the help of the navigation
buttons
.
Confirm with
PRINT key, the instrument is returning
to the menu.
.
Return to weighing mode:
Press the
ESC-key repeatedly until the query
”
SAVE?“ appears.
Confirm query by pressing the
PRINT-key or reject it
by pressing the
ESC-key.
Setup
!
Saveing...
Setup
29.01.08
Setup
29.01.08
Setup
Temperature calibration
Set temp. value [°C]
122
Setup
29.01.08
Setup
Temperature calibration
Set temp. value [°C]
122
Setup
29.01.08
Setup
Temperature calibration
Set temp. value [°C]
122
Setup
29.12.04
Setup
140
139
3 point 00:00