7.1 Adjustment with internal weight after pressing the CAL button

Condition:

Menu setting „Internal adjustm.“, see chap. 7

 In weighing mode press the CAL button, adjustment will take place

automatically.

 After successful adjustment the balance automatically returns to weighing

mode.

 In case of an adjustment error (e.g. objects on the weighing pan) the

display will show an error message „CAL bUt“, repeat adjustment using
the CAL button.

7.2 Automatic adjustment with internal weight

(factory setting)

Condition:

Menu setting „Autom. adjustm.“, see chap. 7

The automatic adjustment with internal weight (see chap. 7.1)

is started

automatically, if the balance

after the weighing balance was disconnected from the mains

after pressing ON/OFF in stand-by mode

after a temperature change of 1.5 °C with non loaded weighing pan / zero

display

after a time interval of 20 min

with non loaded weighing pan / zero display

The automatic adjustment function is always enabled. However you can make a
manual adjustment at any time with the internal or an external weight,
see chap. 7.1 or chap. 7.3.