IMPoRTanT saFeTy InsTRUCTIons

CaUTIon

ReaD anD saVe THese InsTRUCTIons

After unpacking your table, inspect it thoroughly for damage. If you suspect a problem do not use the table and

call our customer service department at 800-916-4612.
Improper use of this device can cause injury. Be sure to read all operating instructions prior to use.
Weight limit (patient and accessories): 227 kg / 500 lbs. distributed evenly. (Bariatric Table only: 340 kg / 750 lbs.)
Do not sit beyond warning line on carbon fiber top. On Lithotripsy / Urology Top - Do not sit on any extensions

(Weight Limit: 200 lbs. / 91 kg.)
The table utilizes four locking casters to permit movement of the table within the imaging suite. Accidental move-

ment of the table may occur. Lock casters prior to accomplishing imaging of the patient. Lock casters when the

patient is mounting & dismounting the table.
The table should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary,

the table should be observed to verify normal operation in the configuration in which it will be used.
The use of Accessories, transducers, and cables other than those specified (with the exception of transducers and cables

sold by the Manufacturer of this device as replacement parts for internal components) may result in increased Emissions

or decreased Immunity of the table.
The table is designed to be a standalone table used with radiographic equipment. This table must not be modified or

incorporated into any other equipment.
There may be pinch points under the table top. Do not have any body part under the top when the table is in

motion.
Be certain that the table is completely lowered prior to discharging an ambulatory patient. The patient may lose balance

and fall. NOTE: The CFPMFXH is a fixed height table. It is recommended that all patients are assisted onto the table

(physical assistance & step stools with side rails is recommended.)
Patient must be secured with Safety Restraint Strap prior to using the tables.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into

service according to the EMC information provided in the specifications at the end of this manual. (Excluding

CFPMFXH)
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment. (Excluding

CFPMFXH)
This table is not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide.
Electrical Shock Hazard. (Excluding CFPMFXH) The power supply/control module is located inside the lifting

column and under the top. No user serviceable parts are inside. Refer servicing to qualified personnel. Unplug wall

connector prior to contact with any cables connected to the power supply.
To avoid the risk of electric shock, this equipment must only be connected an outlet this is properly grounded. See

GROUNDING instructions in this manual.
Grounding reliability can only be achieved when this equipment is connected to an equivalent 3 prong receptacle

marked “Hospital Only” or “Hospital Grade”.
Oakworks medical tables are designed and built to provide many years of dependable service. Please follow local

regulations and laws when disposing of the electrical components used in this table.