Humidity level – Shellab SCO26H-2 User Manual

Page 13

Advertising
background image

4/ is the Over Temp Limit Thermostat set high enough?

– for diagnostics, should

be turned fully CCW with the pilot light never on
5/ has the fuse/circuit breaker blown?

Indicated chamber temperature
unstable

1/ ±0.1 may be normal
2/ is fan working?

– remove top panel and verify movement of cooling fan in

center of shadow box
3/ is ambient radically changing

– either door opening or room airflow from

heaters or air conditioning?

– stabilize ambient conditions

4/ sensor miss-located, damaged or wires may be damaged - check mounts for
control and Over Temp Limit Thermostat sensors, then trace wires or tubing
between sensors and controls
5/ calibration sensitivity

– call Customer Service

6/ high limit set too low

– be sure that Safety is more than 5 degrees over

desired set point; check if Over Temp Limit pilot light is on continuously; turn
controller knob completely clock-wise to see if problem solved then follow
instructions in section 6.4
7/ electrical noise

– remove nearby sources of RFI including motors, arcing

relays or radio transmitters
8/ bad connection on temperature sensor or faulty sensor

– check connectors

for continuity and mechanical soundness while watching display for erratic
behavior; check sensor and wiring for mechanical damage
9/ bad connections or faulty solid state relay

– check connectors for mechanical

soundness and look for corrosion around terminals or signs of arcing or other
visible deterioration
10/ Water jacket empty or low

– check indicator warning light or water level at fill

port in back of unit.
11/ Peltier condensate collector failure

Will not maintain set point

1/ assure that set point is at least 5 degrees over ambient
2/ see if ambient is fluctuating

Display and actual (from reference
thermometer) don’t match

1/ calibration error

– see operator’s manual

2/ temperature sensor failure

– evaluate if pilot light is operating correctly

3/ controller failure

– evaluate if pilot light is operating correctly

4/ allow at least two hours to stabilize
5/ see if reference thermometer is certified

Reference thermometer does not match
digital display

See Temp Calibration in Section 10

Can't adjust set points or calibration

1/ turn entire unit off and on to reset
2/ if repeatedly happens, call Customer Service

Calibrated at one temperature, but not
at another

This can be a normal condition when operating temperature varies widely, e.g.,
>30C° difference between setpoints. For maximum accuracy, calibration should
be done at close to the set point temperature.

HUMIDITY LEVEL

Can't achieve rated humidity/temp

1/ relative humidity sensor or controller failure
2/ check for bad door seal
3/ check for leaking water around steamer
4/ confirm a sufficient and distilled water source
5/ calibrate humidity sensor with independent reference
6/ assure that pressure relief valve is closed
7/ assure that steamer is working (see Humidity Section-4)

Can't adjust set points or calibration

1/ confirm all wire connections
2/ confirm software revision
3/ call Customer Service

Steam generator not working

1/ check if fill solenoid, injection valve, relief valve, float switch and relays are
working

– see schematic in manual

2/ check for water leaks around steamer