PULSER:

By inhibiting heat build-up in the sample, the pulse function enables

safe treatment of temperature sensitive samples at high intensity. In addition,
pulsing enhances processing by allowing the material to settle back under the
probe after each burst. The ON and OFF pulse duration can be set independently
from 01 second to 59 seconds. During the OFF portion of the cycle, the red
indicator on the

PULSE

key will illuminate. If the OFF portion of the cycle

exceeds three seconds, a cautionary message

- Sonics in OFF Cycle -

will warn

the operator against touching the ultrasonic probe. To set the pulser, press

PULSE

key.

The screen will display:

Pulse on

--

sec

Pulse off

--

sec


Using the numeric keys, set the ON portion of the cycle, and press the ENTER/REVIEW
key.

The screen will display:

e.g.

Pulse on 01 sec
Pulse off

--

sec


Using the numeric keys set the OFF portion of the cycle.

The screen will display:

e.g.

Pulse on 01 sec

Pulse off 01 sec

Press the ENTER/REVIEW key.

The screen will display:

Off Cycle

On Cycle

Time 0:30:00
Pulse 01 01 Ample 40

%