Low surface tension liquids - organic solvents, Section 1 – installation, Inspection – Sonics VCX134 User Manual

Page 4: Caution

Advertising
background image

CAUTION

LOW SURFACE TENSION LIQUIDS – ORGANIC SOLVENTS

The probes (solid or with a replaceable tip) are tuned elements that resonate at a specific
frequency. When using a ½” (13 mm) probe with a replaceable tip, if the replaceable tip
is removed or isolated from the rest of the probe, the probe will no longer resonate at that
frequency, and the power supply will fail.
Unlike aqueous (water based) solutions, which rarely cause problems, samples containing
solvents or low surface tension liquids are problematic. These liquids penetrate the
probe/replaceable tip interface, and force the particulates into the threaded section
isolating the tip from the probe.
When processing samples containing solvents or low surface tension liquids

ALWAYS

use a solid probe





SECTION 1 – INSTALLATION

INSPECTION


Prior to installing the Ultrasonic Processor, perform a visual inspection to detect any
evidence of damage, which might have occurred during shipment. Before disposing of
any packaging material, check it carefully for small items.

The Ultrasonic Processor was carefully packed and thoroughly inspected before leaving
our factory. The carrier, upon acceptance of the shipment, assumed responsibility for its
safe delivery. Claims for loss or damage sustained in transit must be submitted to the
carrier.

If damage has occurred, contact your carrier within 48 hours of the delivery date.

DO

NOT OPERATE DAMAGED EQUIPMENT.

Retain all packing materials for future

shipment.







4

Table of Contents

Table of Contents

Advertising
This manual is related to the following products: