TIME:

To set the processing time, press the

TIME

key.


The screen will display:


Using the numeric keys, set the processing time as required:


Press the

ENTER/REVIEW

key. The screen will display:

NOTE

The probe is tuned to vibrate at a specific frequency. If the resonant frequency of
the probe has changed, due to cavitation erosion or fracturing, a minimum reading
will not be obtained. If an overload condition exits, or if minimum reading cannot
be obtained (less than 20%) with the probe out of the sample, check the
instrument without the prove to determine which component might be defective.
If minimum reading is obtained using the converter without the probe, the probe
is defective and should be changed.
A loose probe will usually generate a loud piercing sound.
Refer to Section III if an overload condition exists.

Time Setting

Hrs:__ Min:_ _ Sec:_ _

e.g
.

Time Setting

Hrs: 5 Min: 30 Sec: 25

TIME 5:30:25 TEMP __ __

?

C

PULSE__:__:__ AMPL 40 %