Sonics VCX600 User Manual

Page 20

Advertising
background image

20

To retrieve from memory the parameters stored under that ID number, press the

ENTER/REVIEW

key. The screen will display:














TIME 5:30:25 TEMP 35

?

C

PULSE 2.5 : 1.0 AMPL 40 %

NOTE

To review all the information that has been stored, press keys

0 to 9

one at a

time.

IMPORTANT

Proper care of the probe is essential for dependable operation. The intense cavitation will,
after a prolonged period of time, cause the tip to erode, and the power output to decrease
without showing up on the wattmeter. The smoother and shinier the tip, the more power will
be transmitted into the sample. Any erosion of the probe tip will increase the rate of future
erosion. For that reason it is recommended that after every 5 or 6 hours of use the tip be
examined, and if necessary, polished with emery cloth or an abrasive wheel. Since the probe
is tuned to vibrate at a specific frequency, it is most important that only the contaminated
surface be removed. This procedure can be repeated as long as the wattmeter reads less than
20 watts with the probe out of the sample, when the AMPLITUDE control is set at 100. If
the wattmeter reads over 20 watts the probe or replaceable tip should be replaced with a new
one.

Advertising
This manual is related to the following products: