Sonics VCX600 User Manual

Page 8

Advertising
background image

8











NOTE

Should it become necessary to remove a probe, use the wrenches supplied. If the
probe has been attached to the converter for a long period of time it might be
necessary to use a vise. Be sure the vise has soft jaws or other means to prevent
scratching. Secure the wide diameter portion of the probe in the jaws of the vise.
Never grip the converter in the vise. Using a wrench, twist the converter off the
probe. A tap of a hammer may be applied to the end of the wrench. Never attempt
to remove the probe by twisting the converter housing, as this may damage the
electrical connections within the housing.

Advertising
This manual is related to the following products: