983X / 963X Manometer Calibration Procedure
ESC
LXW
ENTER
SAVE
Range: xx ~ xx psi
(remark2)
Long press
36
EXIT
Cali
PRN
Temp:
P: 0.1psi
ESC
LXW
ZERO
LOW
CALIBRATION
HI
REC
Cali:
PSI
(remak1)
* * *
Press up/down
or number keys to adjust
ESC
LXW
ENTER
SAVE
Cali:
PSI
(remak3)
* * *
Wait
Cali:
PSI
(remak3)
* * *
Range: xx ~ xx psi
Cali:
PSI
(remak4)
* * *
After 5 seconds
OK ( or NG)
EXIT
1.There are up to three calibration points. Please refer to below list:
Models
Cali.(HI)
psi range
Cali. (LOW)
Adjustable Range(psi)
9832/9632
9835/9635
98315/96315
9833/9633
9831/9631
0~+/-2psi
0~+/-5psi
0~+/-15psi
0~+/-30psi
0~+/-100psi
1.600psi
4.000psi
12.00psi
24.00psi
80.00psi
-1.600psi
-4.000psi
-12.00psi
-24.00psi
-80.00psi
1.400~1.800 / -1.400~-1.800
3.500~4.500 / -3.500~-4.500
10.50~13.50 / -10.50~-13.50
21.00~27.00 / -21.00~-27.00
70.00~90.00 / -70.00~-90.00