3 specifications and environmental conditions 9, 1 specifications 9, 2 environmental conditions 9 – ThermoWorks 9135-156 HART SCIENTIFIC 3-POINT IR CALIBRATOR User Manual

Page 11: Specifications and environmental conditions, Specifications, Environmental conditions, 3specifications and environmental conditions, 1 specifications, 2 environmental conditions

Advertising
background image

3

Specifications and Environmental
Conditions

3.1

Specifications

Set-points

50°C (122°F), 100°C (212°F), and 150°C (302°F)

Accuracy

±1.0°C

Stability (2-sigma)

±0.1°C (±0.18°F)

Target Emmisivity

0.95

Target Area

1.5 in. (38 mm) diameter

Test Well Size

0.125 in. (3.18 mm) diameter

Heating Time

25 to 150°C: 3 minutes typical

Stabilization Time

3 minutes typical

Cooling Time

150 to 50°C: 25 minutes typical

Power

115 VAC (±10%), 1.0 A, or 230 VAC (±10%), 0.5 A, 50/60 Hz, 125 W.

Size

1.8" (46 mm) H x 4.4" (112 mm) W x 7.8" (198 mm) D

Weight

1.5 lb. (0.7 kg)

Safety

OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per
IEC1010-1

3.2

Environmental Conditions

Although the instrument has been designed for optimum durability and trou-
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean-
ing recommendations can be found in the Maintenance section of this manual.

The instrument operates safely under the following conditions:

• temperature range: 5-40°C (41-104°F)
• ambient relative humidity: maximum 80% for temperature <31°C, de-

creasing linearly to 50% at 40°C

• pressure: 75kPa-106kPa
• mains voltage within ±10% of nominal
• vibrations in the calibration environment should be minimized
• altitude less than 2,000 meters

9

3 Specifications and Environmental Conditions

Specifications

Advertising