Triner Scale VP-70 Manual User Manual

Page 31

Advertising
background image

Page 6-8

6.2.11 How to perform Scale Calibration











1. Press the F1 key.
2. Press the SCREEN DOWN key.
3. Press the SERVICE soft key.
4. At the password prompt key in (336699) then press ENTER.
5. Press the SCREEN DOWN key.
6. Press the CALIBRATE soft key.

The scale will prompt for the “Latitude Zone”

7. Key in “99” then press ENTER.

Scale prompts to remove all items from the platter

8. Make sure the platform is clear then press the ENTER key.

Scale prompts to place 7 pounds onto the scale

9. Place 7 pounds of precision test weights on the scale platform then press the

ENTER key.

Scale displays “Please Wait” briefly and then prompts you to place 50 pounds
onto the scale

10. Place 50 pounds of precision test weights on the scale platform then press

the ENTER key.

Scale displays “Please Wait” briefly and then indicates “Calibration Saved” if
successful.

Your VP-70 postal scale is FACTORY CALIBRATED. There is no need

to calibrate the scale!


DO NOT attempt to calibrate the scale unless you have been instructed by Triner Scale or one of
their authorized distributors. If you have any questions about the calibration call Triner Scale at
1-800-238-0152.

Advertising