Replacing diaphragms and valves

+

Please read section ”Replacing diaphragms and valves” com-

pletely before starting maintenance.

The pictures may show other versions of pumps. This does not change

the method of replacing diaphragms and valves.

➨

Never operate the pump if covers or other parts of

the pump are disassembled.

➨

Before starting maintenance,

disconnect the electri-

cal power cord. Wait two minutes after isolating the

equipment from AC power to allow the capacitors to

discharge.

➨

Ensure that the pump cannot be operated accidentally.

➨

Note: The pump might be contaminated with the pro-

cess chemicals that have been pumped during opera-

tion. Ensure that the pump is decontaminated before

maintenance.

➨

Avoid the release of pollutants.

+

Never operate a defective or damaged pump.

+

Take adequate precautions to protect people from the

effects of dangerous substances that may have con-

taminated the pump and may be released upon dis-

assembly. Ensure that the maintenance technician is

familiar with the safety procedures which relate to the

products processed by the pumping system.

Use appropriate protective clothing, safety goggles

and protective gloves.

+

Allow sufficient cooling of the pump before starting

maintenance.

+

Vent the pump and isolate it from the vacuum sys-

tem before you start maintenance.