Replac- ing the diaphragm, Above – VACUUBRAND PC 3001 basic pro User Manual

Page 45

Advertising
background image

page 45 of 54

+

Never use a pointed or sharp-edged tool to remove parts (e.g., screw-

driver). We recommend using a rubber mallet or compressed air (to be

blown carefully into port).

For replacing the diaphragm refer to the section

”

Replac-

ing the diaphragm

”

above.

➨

Raise the housing cover.

➨

Pull the tubing off the hose connector at the

the housing cover. Do not kink the tubing.

➨

Remove the head cover (C) carefully from

the housing cover (A). Note position and

alignment of valves (B). Remove the valves.

+

Replace valves if necessary.

+

Use petroleum ether or other industrial sol-

vent to remove deposits. Do not inhale va-

pors.

Assembling the housing cover at the side of the ON/OFF

switch

➨

Put on housing cover (A).

Pay attention to a correct guidance of the connection hose towards the

inlet.

+

Move housing cover slightly to ensure that the head covers are cor-

rectly positioned.

➨

Screw in the Allen head screws at the head

cover in a diagonal pattern with a 4 mm

wide Allen key, loosely at first, to align, then

tighten.

+

Maximum torque: 4.4 ft

.

lb

f

(6 Nm).

Advertising