If the old diaphragm is difficult to separate from the diaphragm support

disc, immerse assembly in naphtha or petroleum ether. Do not inhale

vapors!

D

E

F

E

F

D

➨

Position new diaphragm (E) between dia-

phragm clamping disc with square head

screw (D) and diaphragm support disc (F).

+

Note: Position diaphragm with pale side

towards diaphragm clamping disc (facing

pump chamber).

+

Make sure that the square head screw of

the diaphragm clamping disc is correctly

seated in the guide hole of the diaphragm

support disc.

➨

Lift the diaphragm at the side. Place the di-

aphragm carefully together with diaphragm

clamping disc and diaphragm support disc

in the diaphragm key.

+

Avoid damage of the diaphragm: Do not ex-

cessively bend or crease the diaphragm.

+

Assemble the original number of washers

(G) between diaphragm support disc (F)

and connecting rod (H).

➨

Screw diaphragm clamping disc (D), dia-

phragm (E), diaphragm support disc (F),

and washers (G) to connecting rod (H).

➨

Optimum torque for the diaphragm support disc: 3 ft

.

lb

f

(4 Nm).

+

If necessary, put the pumping unit in its upright position to screw the

diaphragm clamping disc to the connecting rod. Subsequently lay the

pumping unit again on its side.

➨

Bring the diaphragms (E) into a position, in

which they are in contact with the housing

(I) and centered with respect to the bore.

Advertising