Notice – VACUUBRAND PC 301 User Manual

Page 28

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999147 / 17/06/2014

page 28 of 40

A

service manual with exploded view drawings, spare part lists and directions for

repair is available on request (only in German or English).
☞

The service manual is for trained service people.

NOTICE

Fault

Possible cause

Remedy

❑ Pump does not achieve

its ultimate vacuum or

usual pumping speed.

➨ Leak in the pipeline or vacu-

um system?

✔ Check pump directly with

DVR 2 for ultimate vacuum

- close inlet of pumping unit

to do so - then check con-

nection, pipeline and vacuum

system if necessary

➨ Long, narrow line?

✔ Use lines with larger dia-

meter, length as short as

possible.

➨ Pump has been exposed to

condensate?

✔ Allow pump to run for some

minutes with atmospheric

pressure at the inlet.

➨ Deposits have been formed

inside the pump?

✔ Clean and inspect the pump

heads.

➨ Diaphragms or valves dam-

aged?

✔ Replace diaphragms and/or

valves.

➨ Outgassing substances or

vapour generated in the pro-

cess?

✔ Check process parameters.

❑ Pump too noisy.

➨ Atmospheric or high pressure

at the pump inlet?

✔ Connect hose or silencer to

pump outlet.

➨ Diaphragm crack or dia-

phragm clamping disc loose?

✔ Perform maintenance.

➨ Other than above mentioned

causes?

✔ Contact local distributor.

❑ Pump seized.

✔ Contact local distributor.

Advertising