VACUUBRAND PC 301 User Manual

Page 33

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999147 / 17/06/2014

page 33 of 40

➨

Bring diaphragm into a position in which diaphragm is in

contact with housing and centred with respect to bore.

➨

Lay the pumping unit down and support it appropriately.

➨

Assemble head cover and valves.

☞

Check for correct position (see fig. below).

It is essential to mind the position and orientation of the head cov-

ers and the valves!

Scheme of pump head with head covers and valves

valves at the outlet

(round centred opening

under valve)

valves at the

inlet

(kidney-shaped

opening beside

valve)

”side with hose connection”

”motor side”

Advertising