Remove head covers from housing cover and check

valves. Note position of valves and remove.

☞

Replace valves if necessary.

Use petroleum ether or industrial solvent to remove de-

posits. Do not inhale.

☞

Check diaphragm for damage and replace if necessary.

View of the disassembled pump head parts

6

1

7

3

4

5

11

10

8

2

9

1: Housing cover

2: Housing cover insert

3: Valves

4: Head cover

5: Diaphragm clamping disc with

connecting screw

6: Diaphragm

7: Diaphragm support disc

8: Washer

9: Rod

10: Housing

11: Bearing plate