Max. permitted pressure at the pressure transducer: 1.5 bar

(absolute).

The display flashes at a pressure higher than 1080 mbar.

Release pressure immediately.

Risk of bursting!

Inside a vacuum system where evaporation occurs (e.g.

rotary evaporator) the vacuum is not uniform, e. g. a con-

denser acts as pump or the vacuum in the pipeline is lower

than in the system. Therefore carefully choose position

where to connect an

external gauge head (if applicable).

Condensate and deposits in the pressure transducer affect the measuring result.

Clean pressure transducer if necessary, see section “maintenance“.

Connection of components at the controller

At the rear side of the controller are connections for coolant

valve and/or external pressure transducer and/or external

venting valve, a venting connection and a serial interface RS

232C.

- Insert cable of components into the controller. Do not cant

when assembling or removing plug connections!

outlet

(gas!; hose nozzle

10 mm)

Separator at the inlet and exhaust waste vapour condenser

Round bottom flasks:

The catchpot at the inlet prevents droplets and particles from

entering the pump.

Lifetimes of diaphragms and valves are enhanced.

Improves vacuum in case of condensation.

Both round bottom flasks are coated with a protective layer to

prevent disintegration in case of breakage or implosion.

Assemble the catchpots at the inlet and at the outlet using

joint clips.

Exhaust waste vapour condenser:

Assemble hose nozzles for coolant inlet and coolant outlet

pipelines at the exhaust waste vapour condenser.

The

exhaust waste vapour condenser enables an efficient

condensation of the pumped vapours at the outlet.

No backflow of condensates.

Controlled recovery of condensates.

Next to 100% solvent recovery.

The isolation cover protects against glass splinters in case

of breaking, acts as thermal isolation to avoid condensa-

tion of humidity and is intended to absorb shocks.

coolant outlet

(hose nozzle 6-8 mm)

coolant inlet

(hose nozzle 6-8 mm)

overpressure

safety relief

device

catchpot at the

outlet

catchpot at the

inlet

Assembling the hose nozzle with union nut at the inlet:

Take the hose nozzle with attached compression ferrule and union nut out of the

round bottom flask and put onto inlet connection.

Tighten the union nut by hand until you can feel the stop. Then tighten an addi-

tional 1/4 rotation with an open-ended wrench size 17 for final installation.