VACUUBRAND PC 620 NT User Manual

Page 45

Advertising
background image

page 45 of 63

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. Manual-no.: 999188 / 19/05/2009

Fault

Possible cause

Remedy

❑ No digital pressure read-

ing, warning triangle is

flashing, five or seven

blips*.

Five blips: External sensor

defective or removed?

Seven blips: Internal sensor

defective?

Plug in sensor or use new

one or reconfigure without

sensor.

Contact local distributor.

❑ Warning triangle and

pump symbol are flash-

ing, six blips*.

➨ NT VARIO / VARIO-B pump

and VMS** connected?

✔ Disconnect one component,

restart controller.

➨ Fault at the VARIO pump or at

the VMS?

✔ Check pump and/or VMS or

reconfigure.

➨ Fault at the NT VARIO /

VARIO-B pump?

✔ Check pump and cable con-

nections, restart controller.

❑ Clock symbol is flashing,

control has stopped.

➨ Preselected process time has

passed?

✔ Confirm by pressing key

START/STOP.

❑ Internal venting valve

does not react, valve

symbol is displayed.

➨ External pressure transducer

connected and active?

✔ Select internal pressure

transducer or connect exter-

nal venting valve.

❑ Venting valve does not

respond, valve symbol is

displayed.

➨ Venting valve contaminated? ✔ Clean valve.

❑ Function ”Vac control”:

Control stops, ”arrow up”

is flashing.

➨ Preset maximum pressure

exceeded?

✔ Confirm by pressing key

START/STOP. Adapt maxi-

mum pressure value if nec-

essary.

❑ Function ”Pump down”:

Control stops, ”arrow

down” is flashing.

➨ Pressure below preset mini-

mum pressure?

✔ Confirm by pressing key

START/STOP. Adapt mini-

mum pressure value if nec-

essary.

❑ Function is not displayed,

no menu.

➨ No controllable device con-

nected (valve, VMS, VARIO

pump)?

✔ Connect devices or use con-

troller as measuring device.

❑ Controller does not

respond when pressing

keys (except ON/OFF).

PC symbol is displayed.

➨ Controller in remote mode?

✔ Control CVC 3000 via in-

terface or switch off remote

mode.

❑ Controller does not

respond when operat-

ing any keys. No change

after switching off/on.

✔ Contact local distributor.

❑ Pump does not start or

stops immediately. Warn-

ing triangle and pump

symbol are flashing.

➨ Pump has been exposed to

condensate?

✔ Allow pump to run for some

minutes with atmospheric

pressure at the inlet.

➨ Device fuse blown?

✔ Identify cause of failure. Re-

place device fuse.

➨ Overpressure in the exhaust

line?

✔ Open exhaust line, ensure a

free (pressureless) outlet line.