Slot the blinds into the head cover.

Put head cover cowling on.

Slide the head cover cowling in the grooves of the blinds

and under the connection fasteners.

Install the washers. Use a Torx screw driver TX20 to screw

in the 4 screws fixating the head cover cowling.

Slip connecting tube onto hose connection of valve head.

Slide on the tube and the hose clip until touching the nose

at the valve head.

Close hose clip with flat pliers.

Screw in the socket head screws at the head covers di-

agonally first slightly with a hex key size 5, then tighten.

Recommended torque: 12 Nm.

Position clamping bracket with countersunk bores up-

wards.

Align the countersunk bores with the threaded pegs.

Fasten the countersunk screws slightly and correct the

alignment of the valve heads if necessary.

Tighten countersunk screws with Torx screw driver TX20.

Torque: 3 Nm.

Bring the diaphragms into a position in which they are in

contact with the housing and centred with respect to the

bore.

Put on head cover with valve heads and connections.

Pay attention to the correct orientation of the head covers:

Housing with cylinder pin: The cylinder pin at the pump

housing has to fit into the recess at the head cover.

Housing with mark: Align the recess at the head cover with

the mark at the pump housing.