Keep the control line away from heated surfaces.

• Provide a firm, level platform for the equipment. Check

that the system which you are going to evacuate is

mechanically stable. Check that all fittings are secure.

Ensure a stable position of the pump without any me-

chanical contact other than the pump feet.

• Comply with

maximum permissible pressures at in-

let and outlet and with

maximum permissible pres-

sure differences between inlet and outlet. See section

„Technical data“, pg. 22. Do not operate the pump

with overpressure at the inlet.

• Avoid overpressure of more than 17.5 psi absolute (1.2

bar absolute) in the event that gas or inert gas is con-

nected to the pump, or to a venting valve.

Note: Flexible elements will shrink when evacuated.

• Connect hoses gas tight at inlet and outlet of the pump.

• Ensure that no foreign objects can be drawn into the

pump.

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

Make sure ventilation is adequate to maintain recommend-

ed operating temperature. Keep a minimum distance of

2 in (5 cm) between the cooling fan and surrounding items

(e.g., housing, walls, etc.), or else install an external auto-

matic ventilation system. Check fan regularly for dust/dirt.

Clean fan guard grill if necessary to avoid a reduction of

ventilation.

Use only hoses at the inlet and outlet of the pump with

an inner diameter at least as large as the diameter of the

pump’s tubing (to avoid overpressure at the outlet, and

reduction of pumping speed at the inlet).