Make sure that the exhaust pipeline cannot become

blocked. If there is an exhaust isolation valve, make

sure that you cannot operate the equipment with the

valve closed to avoid a

risk of bursting!

+

Always provide a free and pressureless exhaust outlet

to avoid damage to pump valves and risk of bursting.

+

Comply with applicable regulations when disposing of

chemicals. Take into consideration that chemicals may

be contaminated. Take adequate precautions to pro-

tect people from the effects of dangerous substances

(chemicals, thermal decomposition products of fluoro-

elastomers). Use appropriate protective clothing and

safety goggles.

+

Interruption of the pump (e.g., due to power failure),

failure of connected components or of parts of the sup-

ply, or change in parameters must not be allowed to

lead to dangerous conditions. In case of a diaphragm

failure or in case of a leak in the manifold, pumped

substances might be released into the environment or

into the pump housing or motor.

Comply with all notes regarding proper use of the

pumps, as well as operation and maintenance guid-

ance.

+

The residual

leak rate of the equipment might render

possible an exchange of gas, albeit extremely slight,

between the environment and the vacuum system.

Adopt suitable measures to prevent contamination of

the pumped substances or the environment.

• Ensure that no parts of your clothing, hair or fingers

can be caught or drawn in at the inlet of the pump.

Never insert fingers or drop any other object into the

inlet or outlet.

• Pay attention to the safety symbol ”hot surfaces” on

the equipment. Hot parts may cause burns if touched.