Service kit for MD 4 VARIO-SP (diaphragms, valves, O-rings)....................696834

Service kit for MZ 2 VARIO-SP (diaphragms, valves) ..................................696835

O-ring 28 x 1.5 (4x for MZ 2 VARIO-SP) .....................................................637090

Diaphragm key (w/f 66) ..............................................................................636554

The valves and diaphragms are wear parts. If the rated

ultimate vacuum is no longer achieved or in case of in-

creased noise level, the pump interior, the diaphragms

and the valves must be cleaned and the diaphragms and

valves must be checked for cracks or other damage.

All bearings are encapsulated and are filled with long-life

lubricant. Under normal operating conditions, the drive

system is maintenance free.

In demanding circumstances, it may be efficient to check

and clean the pump heads on a regular basis. In normal

use, the lifetime of the diaphragms and valves is typically

15,000 operating hours (at 1500 rpm). Operating the pump

at reduced speed will increase the lifetime accordingly.

- Prevent internal condensation, transfer of liquids or

dust. The diaphragms and valves will be damaged if

liquid is pumped in significant amount.

- Carry out maintenance frequently if the pump is ex-

posed to corrosive media or in case of deposits.

- Regular maintenance will improve the lifetime of the

pump and also protect both users and the environment.

Tools required (metric):
- Diaphragm key width 66 mm

- Open end wrench width 17 / 20 mm

- 5 mm wide Allen key