Installation, General precautions, Mounting the detector on the gc – VICI D-4-I-TQ-R User Manual

Page 10: Pulser module installation

Advertising
background image

6

Installation

General Precautions

Do not use plastic/polymer or copper tubes for gas handling and
interconnectons. Use only stainless steel tubing with Valco gold-plated
ferrules.

Do not turn the unit on until the helium discharge gas is flowing through
the detector.

Do not shut off or disconnect the discharge gas when the detector is hot,
even if the unit is turned off. Turn off the discharge power switch and
allow the detector to cool down naturally before disconnecting or shutting
off the discharge gas.

Do not cover the unit with materials or devices which would restrict air
circulation.

Mounting the Detector on the GC

1. Remove the FID flame tip.

2. Install the PDD on the FID heater block. Make sure you interpose the

silver seal.

3. Tighten the nut to secure the detector.

Pulser Module Installation

1. Mount the pulser module on the GC as shown below.

2. Connect +24VDC power to the pulser module from the GC.

PDD C

ARD

puls

ed d

isch

arg

e m

odu

le

BIAS

TO PNEUMATIC
MODULE

PULSED DISCHARGE

MODULE

PDD CONTROL

CARD

DETECTOR
MOUNTING NUT

ELECTROMETER

GAS
INLET

Figure 3: Mounting the detector on the GC