Capillary column connection, Packed column connection – VICI D-4-I-TQ-R User Manual

Page 15

Advertising
background image

11

Installation

Capillary Column Connection

If the capillary column adapter is installed in the column inlet:

1. Make a mark on the column 14.5 cm from the end.

2. Remove the knurled nut and plug from the capillary column adapter in the

column inlet at the bottom of the detector. Slide the nut over the end of
the column, followed by the appropriate column ferrule (FS.4 or FS.5, or
ZF.5V for megabore).

3. Seat the ferrule in the detail of the column adapter and begin sliding the

column through the capillary column adapter and into the column inlet.

4. Get the nut started on the threads and tighten it until you feel it contact

the ferrule, then back off half a turn. Slide the column into the column
inlet until the mark is flush with the surface of the knurled nut, and secure
the column in the adapter by tightening the knurled nut

finger tight only.

If the capillary column adapter has been removed, reinstall it:

1. Unscrew the liner as far as it will go, then screw the fitting body into the

column inlet fingertight.

2. While using a 1/8" wrench to prevent rotation of the liner (the part with the

seat for the column ferrule), use a 1/4" wrench to tighten the body of the
adapter until the ferrule has sealed. The liner

will deform if it rotates.

3. Proceed to Step 1 above.

Packed Column Connection

To prevent detector contamination, we strongly
recommend disconnecting the column from the
detector during column bakeout procedures.

The D4 is optimized for packed columns. The column tubing must be
thoroughly cleaned and baked before the column is packed. Even when
the best care is taken in column tubing cleaning and in the support and
stationary phase selection, a new column will often bleed compounds,
resulting in a considerable increase in the detector baseline. This initial
bleed will usually be reduced to acceptable levels after the column is
conditioned with clean carrier gas flow for several hours at the
recommended bakeout temperature.

1. Loosen and remove the knurled nut and plug of the capillary column

adapter, (or remove the column ferrule and the column if one has been
installed).

2. Use a 1/8" wrench to hold the liner – that part of the adapter in which the

column ferrule sits. While the 1/8" wrench keeps the liner from rotating,
use a 1/4" wrench on the fitting body to loosen the adapter 1/2 turn.

14.5 cm

/

COLUMN
INLET

COLUMN
FERRULE

NUT

CAPILLARY
COLUMN

CAPILLARY
COLUMN
ADAPTER
(IZERA1.5)

13 cm

/

COLUMN
INLET

PACKED
COLUMN
ADAPTER

PACKED

COLUMN

(I-23642-TQ)