2 buffer sets for calibration, See section 10.1.2 b, Uffer – YSI MultiLab 4010-3 User Manual
Page 86: Sets, Calibration

Settings
MultiLab 4010-3
86
ba76143e03
08/2013
10.1.2 Buffer sets for calibration
You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.
Reset
-
Resets all sensor settings to the delivery condition (see
section 10.7.1 R
Menu item
Possible
setting
Explanation
No.
Buffer set *
pH values
at
1
YSI *
4.000
7.000
10.000
25 °C
2
ConCal
Any
Any
3
NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers
1.679
4.006
6.865
9.180
12.454
25 °C
4
TEC
Technical buffers
2.000
4.010
7.000
10.011
25 °C
5
Merck 1*
4.000
7.000
9.000
20 °C
6
Merck 2 *
1.000
6.000
8.000
13.000
20 °C
7
Merck 3 *
4.660
6.880
9.220
20 °C
8
Merck 4 *
2.000
4.000
7.000
10.000
20 °C
9
Merck 5 *
4.010
7.000
10.000
25 °C