YSI 610DM User Manual

Page 32

Advertising
background image

YSI Incorporated

Displays and Loggers

30

3.

Set the PC to run from the floppy drive which holds the YSI programs (type A: or B:
on the PC, and press Enter on the PC).

4.

Disconnect the wall charger from the 610 if it is connected.

5. Turn off the 610 by pressing Power, or by pressing Esc then Power. Reset the 610 by

tapping the Esc and Skip keys simultaneously, and then releasing them immediately.
If you did this correctly the message RESET DONE or MASTER RESET will
appear in the 610 display. If you do not get either message, go back to step 5.

6.

Type the following 5 keys on the 610: D C ? Del Bksp. Verify you see PMODE
on the 610 screen. If you make a typo, press Skip to backspace over the typo and
retype it. Once you see PMODE press the Z key. The 610 screen displays
PROGRAMMING MODE !!LOADING!!

7.

On the PC at the DOS prompt, type the command "LOAD4 -ROM -F128" and press
Enter (also on the PC). This runs the transfer program. Loading begins at once and
takes about 5 minutes. Loading is complete when the DOS prompt returns to the PC
screen, and you hear a chirp from the 610.

8.

Verify the message ALL ROM BANKS SENT is on the PC screen, and the word
COMPLETE has appeared on the 610 screen. If either message is not shown, go back
to step 1.

9.

If the 610 has never been loaded with YSI software before, you need to do items A-E
below. These items are shown here for completeness. Otherwise proceed to step 10.

10.

(A) Type the 5 keys: ' Bksp _ G D. Verify you see SETUP on the 610 screen.
Press the Z key.

(B) Press Q until baud rate is 9600, then press "
(C) Press Q until word format is 8D+1 STOP then press "
(D) If bank number is not 0, press Q Y Z then press "
(E) If prog addr is not 16384, press Q S P U J N Z then press "

11.

Type the following 3 keys on the 610: F G ". Verify you see RUN on the 610
screen. If you make a typo, press Skip to backspace over the typo and retype it. Once
you see RUN press the Z key. The 610 displays YSI SOFTWARE INITIALIZED for
3 seconds, then shuts itself off.

12.

The 610 is now ready for use.

SECTION 8 WARRANTY AND SERVICE INFO

The dispay/loggers are warranted for two years against defects in workmanship and
materials when used for its intended purposes and maintained according to instructions.
All cables are warranted for one year. Damage due to accidents, misuse, tampering, or
failure to perform prescribed maintenance is not covered. The warranty period for

Advertising