1.3 POWERING UP

After the YSI 610 is shipped, the batteries may need a few minutes of charging before the
unit can be turned on. Normally, whenever the AC adapter is plugged in, the YSI 610
turns itself on. To turn the YSI 610 off when the AC adapter is plugged in, press the
Power button, wait for the power up display to appear and then press the Power button
again. Do not allow the YSI 610 to charge more than 48 hours continuously.

With a sonde plugged in, the YSI 610 powers up into the Run mode and displays
readings.

The message, “No Sonde in Use” may appear occasionally while the YSI 610 is trying to
establish communication to a sonde. If the message does not go away after 60 seconds
check the cables, power and baud rate selection on both the YSI 610 and the sonde to
make sure that they are working properly, and that the baud rates agree (See Section 2,
Sondes
of the YSI 6-Series Operations Manual for information about the proper baud
rate).