YSI 85 User Manual

Page 38

Advertising
background image

Troubleshooting

Section

12

YSI, Incorporated

Model 85

34


SYMPTOM

POSSIBLE

CAUSE

ACTION

12. Main display reads “LErr”

A. In temperature compensated

conductivity mode, temperature
exceeds the values computed using
user defined temperature coefficient
and/or reference temperature.

B. In cell constant cal mode, temperature

exceeds the values computed using
user defined temperature coefficient
and/or reference temperature.

A. & B. Adjust user defined tempco or

reference temperature. (pg. 10)

13. Main display reads “Err”
(Secondary display reads “ra”)

A. System has failed its RAM test check

procedure.

A. Turn instrument OFF and back ON

again.

B. Return the system for service (pg. 26)

14. Main display reads “Err”
(Secondary display reads “ro”)

A. System has failed its ROM test check

procedure.

A. Turn instrument OFF and back ON

again.

B. Return the system for service (pg. 26)

15. Secondary display reads “rEr”

A. Temperature jumper is set to °F and

reading is >199.9°F but <203°F.

A. Return the system for service. (pg. 26)

16. Main display reads “FAIL”
(Secondary display reads “eep”)

A. EEPROM has failed to respond in

time.

A. Return the system for service. (pg. 26)

17. Readings on main display don’t
change

A. Meter is in recall mode.

A. Press MODE button to return to Normal

Operation (pg. 12)