YSI 95 User Manual

Page 18

Advertising
background image

YSI Incorporated

Model 95

14

SECTION 6

SECTION 6

SECTION 6

SECTION 6

MAINTENANCE OF THE MEA SENSOR

MAINTENANCE OF THE MEA SENSOR

MAINTENANCE OF THE MEA SENSOR

MAINTENANCE OF THE MEA SENSOR

6.1

6.1

6.1

6.1 ANODE SERVICE

ANODE SERVICE

ANODE SERVICE

ANODE SERVICE

Warning: Under no circumstances should ammonium hydroxide be used to clean the silver
anode. Ammonium hydroxide will permanently damage the condition of the MEA surface.

The MEA oxygen sensor is, in principle, the same as the conventional Clark oxygen sensor in that
the sensor is made of a silver anode and a gold cathode, but the cathode is a microelectrode array.
Since the current of the MEA oxygen sensor is so much smaller (on average 100 times smaller) than
the current of YSI conventional oxygen sensors, consumption of the silver anode, due to the
formation of AgCl, is minimal during the lifetime of the probe. There should not be any significant
build-up of silver chloride at the surface of the anode for 3 to 4 years, therefore, the anode should
not require chemical cleaning. However, if the surface of the silver anode has become fouled, gently
wet sand it using 400 grit wet/dry sandpaper, rinse thoroughly with deionized or distilled water and
wipe with a wet paper towel until the dark layer is removed. The directions are as follows:

Anode Cleaning Procedures (See figure 6)

ü Rinse the sensor thoroughly after removing the membrane cap.
ü Use wet 400 grit sandpaper to sand away the top layer of the anode by wrapping the sandpaper

around the anode and gently rotating it until the dark layer is removed.

ü Rinse the anode thoroughly with deionized or distilled water and wipe with a wet paper towel.
ü Rinse the anode again with deionized or distilled water.

Cathode

Anode

Press lightly against
sensor surface

Wet

microcloth

Buffing Tool

Figure 6

6.2

6.2

6.2

6.2 CATHODE SERVICE

CATHODE SERVICE

CATHODE SERVICE

CATHODE SERVICE

Warning: Under no circumstances should the gold cathode surface (the MEA surface) be
sanded. Sanding will permanently damage the condition of the MEA surface.

If the MEA oxygen sensor exhibits erratic behavior, such as a current rise at a rate of 1%/hour or
very jumpy readings, it can be serviced by buffing. You can use a few light twists against the
surface with the wet microcloth mounted on the buffing tool provided in the 9503 reconditioning
kit.

Advertising