System : buzzer <boolean, System : buzzer, System : error – Anritsu MP1777A User Manual
Page 105: System : version
Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Section 8 Device Message Detaiis
: SYSTem : BUZZer <boolean>
Command type
Parameter
Function
Example of use
Sequential command
<boolean> = <BOOLEAN PROGRAM DATA>
OFF or 0........Buzzer OFF
ON or 1......... Buzzer ON
Switches the buzzer ON and OFF.
Switches the buzzer ON.
< : SYSTem : BUZZer ON
: SYSTem : BUZZer?
Command type
Response
Function
Example of use
Sequential command
<NR1 NUMERIC RESPONSE DATA>
0 ....................Buzzer OFF
1 ................... Buzzer ON
Makes a query of the buzzer setting status.
> : SYSTem : BUZZer?
<
1
SYSTem : ERRor?
Command type
Response
Example of use
Sequential command
<error/event_number>. <error/event_description>
<error/event_number> = <NR1 NUMERIC RESPONSE DATA>
-32,768 to 32,767
<error/evenUdescription> = <STRING RESPONSE DATA>
> : SYSTem : ERRor?
< -102, "Syntax error"
: SYSTem : VERSion?
Command type
Response
Function
Example of use
Sequential command
<version>
<version> = <NR2 NUMERIC RESPONSE DATA>
YYYY.V.......YYYY (year), V (revision number)
Makes a query of the version of SCPI to which this device conforms.
> : SYSTem : VERSion?
< 1993.0
8>28