B.2 Initializing Message

B.2 Initializing Message

DCL, SDC:

Initializes the message exchange among all the devices on GPIB or only specified devices. When a

GPIB option is mounted.

The following table shows target items of message exchange initialization.

Table B-3

Items of Initialization Target of Message Exchange

Item

Effects of Initalization

Input and output buffers

Input and output buffers will be cleared.

Syntax analyzer, run time controller and response production module

The syntax analyzer, run time controller, and response production module will be

reset.

' Device command including *RST

All the commands to interfere with the execution of these commands will be

cleared.

” Program message to the parameter

All the commands and queries of the Section, where execution is postponed for

the parameter, will be discarded.

~ Processing of the =»^OPC command

Switches the device into the OCIS (Operation Complete Command Idle State).

As a result, operation completion bit will not be set in the standard event status

register.

Processing of the OPC? query

Switches the device into the OQIS (Operation Complete Query Idle State). As a

result, the operation completion bit 1 cannot be set in the output queue.

The MAV bit will be cleared.

Device function

Sections related to message exchange are all switched into the idle state. The

device waits for a message from the controller.