Operation Theorem • 25
Once the operation starts and has received a Trigger In signal, the switch 
module updates the relay pattern to that specified in the first entry of scanlist. 
T
S
is the default debounce time for a switch module, i.e. 5ms for PXI-7931. T
An
is the user specified scan delay time in the scanlist entry, indicating the time 
between the relay being debounced and the exact moment that a measurement 
device takes a new measurement. The actual delay time would be the greater 
of the two times, to guarantee that measurement devices take measurements 
after the signal path is fully settled, and the relays switch as close as possible to 
their maximum operating speed. 
As the scan delay time elapses, the switch module generates Scanner 
Advanced
signal to inform the measurement device to take a new
measurement.
After the measurement completes, the measurement device will generate 
another Trigger In signal to have the switch module update the relay pattern to 
that specified in the second entry of scanlist. 
The handshaking process will continue, until it reaches the end of the scanlist 
(if one-time scanning mode is selected), or when a software scan-abort 
command is received. 
Connecting, Trigger and Synchronize with External DMM 
 
In this example, Agilent
®
33401A 6-1/2 Digital Multimeter (DMM) is used to
demonstrate signal connection for handshaking operation.
The DMM provides two terminals on the rear panel for the handshaking 
operation process, Trig In and VM Comp. The Trig In connects to the Scanner 
Advanced
output on switch module, while the VM Comp to Trigger In. If the
terminal board is used, wire Trig In to TRG_IN on terminal board, and the VM 
Comp
to S_ADV.
Follow the instructions below to ensure the handshake functions properly:
1. Configure ADLINK PXI switch module’s Trigger In to rising-edge
triggered, Scanner Advanced output in active-low pulsating 
mode with pulse width of at least 2us. 
 
2. Configure the DMM to wait for external Trig In before a 
measurement, and generate VM Comp after a measurement. 
Arm the DMM to wait for the first trigger. 
3. Setup the scanlist and auto-scan mode. The first entry in the
scanlist should be set disable wait for the Trigger In, but enable 
Scanner Advanced
output. Succeeding entries should enable
both Trigger In and Scanner Advanced output. Download the