12 s – ENMET PGD2 User Manual

Page 25

Advertising
background image

ENMET Corporation

PGD2

22

12

S

ENSORS

12.1

Flammable Sensors

The PGD2 instrument can be fitted with sensors for the detection of various types of flammable gases. The sensors used are of the
catalytic oxidation type generally referred to as “pellistors”. Sensors of this type consist of a matched pair of elements known as the
detector and compensator.

Either one or two pairs of pellistors may be fitted within the instrument depending upon whether it is required to measure only up to the
Lower Explosive Limit of the gas (one pair fitted) or up to 100% volume of the gas (two pairs fitted). The warm up time for either type is
less than 15 seconds, this takes place when the instrument is switched on. No readings are displayed during this period.

The flammable sensors are non-selective that is, they will detect a range of hydrocarbon gases. However, during manufacture of the
PGD2 instrument the sensors are calibrated to methane gas and the user may then select other gases from the menu selection as
described in Section 6.11. The instrument automatically applies the appropriate calibration factor according to the gas selected so that
the correct reading is given on the display.

12.1.1

LEL Flammable Sensors

Where only one set of pellistors is fitted, to enable measurement up to the Lower Explosive Limit of the flammable gases
(approximately 5% volume), the pellistor detector and compensator are encapsulated into individual stainless steel housings behind
individual stainless steel sintered flame arrestors. The stainless steel housings are contained within in a blue coloured plastic
moulding which also houses the oxygen sensor (if fitted).

Where fitted, the flammable sensor and oxygen sensors are located behind the instruments’ right hand side gas inlet as viewed from
the front.

There are two types of LEL sensors available:-

i). Type SS10 – for general hydrocarbons e.g. Methane, Propane etc. and Hydrogen.

ii). Type VQ41 – for detection of fuels e.g. diesel, aviation fuels etc.


LEL Flammable Sensor Specification

Gases detected

-

Various combustible gases and vapors


Range

-

0 - 100% LEL


Resolution

-

1% LEL


Operating Voltage

-

2 ± 0.1 Volts dc


Operating Current

-

180 mA


T

90

Response Time

-

< 10 Seconds


Linearity

-

Linear 0-100% LEL Methane


Long Term Sensitivity
Drift

-

< 5% signal / month


Long Term Zero Drift

-

< 2% LEL methane / month

Advertising