1 specification, Ntroduction – ENMET PGD2 User Manual

Page 6

Advertising
background image

ENMET Corporation

PGD2

3

5

I

NTRODUCTION

This Instruction Manual No.1 provides information on the General Operation of the PGD2 instrument.
Instruction Manual No.2 The Data Processing-Calibration-Configuration & Programming of the PGD2 provides information on the use
of the specially designed Windows based software package available for use with the instrument.

5.1

Specification

Size

185mm high x 108mm wide x 50mm deep.(excluding boot)

Weight

1Kg.

No. of Gases

Up to four.

Gas Types

Flammable, Toxic, Oxygen (See section 12 for sensor specifications).

Internal Pump

Available to order (suffix P).

3.5 Volts dc.

Power Sources

i). rechargeable battery pack Model PGD/BATT.

Minimum operating times

12 hours.

Visual indications

i). 2 x 16 character LCD display with backlight.
ii). large flashing red lens section used to indicate alarm conditions.
iii). Green ‘Power’ and amber ‘Fault’ indicators in lens.

Audible indication

Electromagnetic alarm in red lens section.

Alarm levels

Flammable : single level instant alarm.
Toxic (each sensor) : 3 levels - instant, STEL, TWA.
Oxygen : 2 levels - instant low, instant high.
Automatic calculation of STEL and TWA readings.

Data Logging Modes

i). Automatic variable rate according to gas levels.

ii). Fixed rate.
iii). Fixed fast rate.
iv). Off.

Display formats

ppm, %, LEL.

Controls

i). Front panel - on/off, right arrow, left arrow, on/off/menu.
ii). Top panel - display backlight, alarm mute.

Temperature range

i). operating -5 to + 40

O

C.

ii). storage -20 to + 50

O

C.

Environmental Rating

IP66.

Humidity range

15% to 95% relative humidity, non-condensing. Operation is possible in
the range 0-100% R.H. however, sensor life will be reduced by
continuous operation at extremes of humidity and temperature.

Electromagnetic compatibility

CE Marked, Tested to EN50081-2, EN50082-1, EN55022 (Limit B).

Air Pressure Operating Range

920 mbar to 1150 mbar.

Advertising