5.0 Maintenance of the EX-5175-EC

C

AUTION

:

Do not open the

EX-5175-EC

S/T

in a classified area.

C

AUTION

:

Do Not Attempt A Span Procedure Without Calibration Gas Applied to The Sensor; if this is done, the S/T is forced

into a calibration fault mode.

Magnetic switches control the

M

ENU

and

S

ELECT

functions. The

M

ENU

and

S

ELECT

switch locations are indicated on the

display panel, see Figure 3. The

M

ENU

switch is used to display the various menu options and make incremental changes to

numbers such as alarm points, calibrations gas, etc. The

S

ELECT

switch is used to select that option, set zero or span digit.

Most maintenance functions are controlled by simple taps of the supplied magnet on the transmitter glass, below the

M

ENU

and

S

ELECT

boxes on the front panel.

5.1 Maintenance Menu

To enter the maintenance menu hold the magnet over the

M

ENU

switch for 2 to 4 seconds

Table 1 indicates the maintenance menu sequence see Figure 5 for a detailed maintenance menu flow chart.

Table 1: EX-5175-EC maintenance Menus Sequence

Example of Display

Function


Normal Display Mode

Measurement of target gas

Hold the magnet over

M

ENU

switch for 2 – 5 seconds to enter the Maintenance Menu

The Power/Fault LED will flash Green – Red to indicate the

EX-5175-EC

is in Maintenance Mode

To exit the maintenance Menu and return to the Normal
Display Mode:

If intended function Tap the magnet over

S

ELECT

switch

Tap the magnet over the

M

ENU

switch to advance to the Zero procedure

For adjusting Zero:

If intended function Tap the magnet over

S

ELECT

switch

N

OTE

:

Not applicable for Oxygen S/T.

Tap the magnet over the

M

ENU

switch to advance to the Span procedure

For adjusting the Span:
If intended function Tap the magnet over

S

ELECT

switch

Tap the magnet over the

M

ENU

switch to advance to each Alarm set point procedures

For adjusting the Alarm 1, 2 and 3 set points:
If Intended function Tap the magnet over

S

ELECT

switch

Tap the magnet over the

M

ENU

switch to advance the mA Span set point procedure

For adjusting the mA Span set point:
If intended function Tap the magnet over

S

ELECT

switch

Taping the

M

ENU

switch without taping the

S

ELECT

switch will allow you to cycle through the menu options.

You must Tap the

S

ELECT

switch in order to change the desired operation.

N

OTE

:

If the S/T fails to respond, the magnet may have become weak and may need to be replaced.

5ppm

Exit

Zero

Span

mA Span

Alarm1

Alarm2

Alarm3