FPM-300/FLS-300/FOT-300

10

Defining a List of Favorite Wavelengths

You must put the wavelengths you want to use on a list of favorite wavelengths (the
FAV list). Only wavelengths on this list are available for measurements.
At the factory (or after recalibration), the list contains 10 calibrated wavelengths.

Note:

The list must always contain at least one wavelength.

To add wavelengths to the FAV list (or to remove them):
1.
Hold down

(power meter)

a few seconds. The unit

enters the FAV list and displays the current
wavelength.

2. Press

(power meter)

to switch between available

wavelengths. An asterisk (

*) appears beside

wavelengths already on the list.

3. Press

to include/exclude the displayed wavelength.

4. Repeat steps 2 to 3 for other wavelengths as necessary.
5. Press

to return to normal mode. If your list is empty, the unit beeps and you

cannot exit the FAV list.