Maintenance
High-Speed Power Meter
93
Cleaning EUI/EUA Adapters
3. Slowly insert the cleaning tip into the adapter until it comes out on the
other side (a slow clockwise rotating movement may help).
4. Gently turn the cleaning tip one full turn.
5. Continue to turn as you withdraw the cleaning tip.
6. Perform steps 3 to 5 again, but this time with a dry cleaning tip (2.5 mm
tip provided by EXFO).
Note: Make sure that you do not touch the soft end of the cleaning tip, and verify
the cleanliness of the cotton tip.
7. Throw out the cleaning tips after one use.