Programming the analyzer, Setting restore – Hanna Instruments PCA 310 User Manual
Page 32

32
Pressing “RIGHT” or “LEFT” arrow keys, another parameter
can be set.
Pressing “ESC” key will return to menu.
Note:
If the edited value is outside the allowed range, a warning
panel appears when “CFM” is pressed. This panel contains
the parameter limits. Pressing again “CFM” or “ESC” will
return to the edit mode.
PROGRAMMING THE ANALYZER
To set the parameters press “MENU” while in the measure
mode. The main Menu is displayed.
Select the appropriate entry as described in next pages.
SETTING RESTORE
The settings are stored in a nonvolatile EEPROM memory. If a
power failure appears the settings are restored after power on.
When power is first time applied to PCA 310-330 analyzers,
the settings are set to factory default values.
At startup the EEPROM content is analyzed and, if some set-
tings are corrupted, a restore procedure is started. The instru-
ment displays the following screen:
Press “CFM” to restore the settings that are out of range. In
this case, the corrupted settings are set to default. In this case
all settings should be checked in to see the values.
Press “SET” to restore to default of all settings.
In extreme situation when the LCD is blank at startup, is pos-
sible to reset the instrument by keeping the keys “UP” + “SET”
+ “MENU” pressed while the instrument is powered on.